We are not the manufacturer, authorised distributor or authorised representative of the Keyence brand in Australia. The private labels and part numbers shown on this website are the property of their respective owners.
If you require FD-Q10C from manufacturer Keyence, please contact FAMAGA.
For more than 9 years our specialists supply spare parts for the repairs and maintenance different kinds of technical equipment from all parts of the planet. Over 750 orders from clients from countries of near and far abroad are in progress daily. Developed partner network provides an opportunity to execute without delay order for Keyence FD-Q10C, while the customer does not have to go over budget.
Geography of deliveries involves Europe, Asia, America and Africa. Australia, India, Indonesia, United States, Iran, Belize, Chile, Egypt, Belgium, South Korea, Bahrain, Bangladesh, Paraguay, El Salvador, Macedonia, Nicaragua, Guatemala, Sierra Leone, Angola, Zambia, Iraq, Cambodia, Mauritania, Uzbekistan, Liechtenstein, Poland, Libya, Armenia, Suriname, Kiribati, Dominica, Venezuela, China, Montenegro, Tanzania, Madagascar, Burundi, Liberia, Myanmar, Italy, South Sudan, Philippines, Togo, Ethiopia, Cape Verde, Lesotho, Benin, Honduras, Papua New Guinea, Kazakhstan, United Arab Emirates, Netherlands There is no restrictions on the size and mass of the cargo. We submit for review different delivery options so that a buyer can choose optimal compromise between cost and delivery time.
Company`s main postulate is transparency, which involves every segment of activity. Clients get only high-quality products, to which relevant documentation from the manufacturer is available. All repair parts and equipment are original, risk of presence of counterfeits are excluded. For permanent partners there are bonus programs and they are valid special pricing.
Become a partner from FAMAGA, get equipment with no delays and at attractive price. Fill in the order form on the website. Our employee will contact you to specify details. Requests are also accepted by phone.
Model | FD-Q10C | |
Supported pipe diameter | 1/4' (8 A), o13 to o16 (mm) 3/8' (10 A), o16 to o18 (mm) | |
Supported pipe materials | Metal pipe/Resin pipe*1 | |
Supported fluids | Various liquid [i.e. water (including DI), oils, chemicals, etc.]*1 | |
Supported fluid temperature | 0 to 85°C (No freezing on the pipe surface)*2 | |
Maximum rated flow | 1/4' (8 A): 20 L/min 3/8' (10 A): 30 L/min | |
Zero cut flow rate | 1.0 L/min (Default)*3 | |
Display method | Status indicator, output indicator, dual row display with 4-digit, 7 segment LED, stability level indicator | |
Display update cycle | Approx. 3 Hz | |
Display resolution | 0.01/0.1/1 (Default : 0.1) (L/min) | |
Response time | 0.5 s / 1.0 s / 2.5 s / 5 s / 10 s / 30 s / 60 s | |
Repeatability | 0.5 s : ±2.0%, 1 s:±1.5%, 2.5 s : ±1.0%, 5 s : ±0.5%, 10 s : ±0.35%, 30 s : ±0.2%, 60 s : ±0.15% (/F.S.*4 (Specific to selected response time)) | |
Hysteresis | Variable | |
Integrated flow unit display | 0.1/ 1/ 10/ 100/ 1000 (Default : 1) (L) | |
Integrated flow data storage cycle | Save to memory every 10 seconds | |
Memory backup | EEPROM (Data storage length: 10 years or longer, Data read/write frequency: 1 million times or more) | |
Power I/O connector | M12 4-pin connector | |
Input/Output (Selectable) | Output (ch.1/ch.2) | Control output/ Pulse output/ Error output (Selectable, Default : ch.1 control output/ ch.2 not used), NPN/PNP setting switchable,open collector output 30 V or less, max. 100 mA/ch., residual voltage 2.5 V or less*5 |
Analogue output (ch.1/ch.2) | 4 to 20 mA / 0 to 20 mA (Selectable, Default: not used), load resistance 500 ? or less*5 | |
External input (ch.2) | Integrated flow reset input/ Flow rate zero input/ Origin adjustment input (Selectable, Default: not used), short-circuit current 1.5 mA or less, input time 20 ms or more*5 | |
Power source | Power supply voltage | 20 to 30 VDC , ripple (P-P) 10% max, Class2/LPS |
Current consumption | 100 mA or less (Load current excluded)*6 | |
Protection circuit | Power supply reverse connection protection, power supply surge protection, each output short-circuit protection, each output surge protection | |
Environmental resistance | Enclosure rating | IP65/IP67 (IEC60529) |
Ambient temperature | -10 to +60 °C (No freezing) | |
Relative humidity | 35 to 85 % RH (No condensation) | |
Vibration resistance | 10 to 55 Hz, Double amplitude 1.5 mm, 2 hours in each of the X, Y, and Z directions | |
Shock resistance | 100 m/s2, 16 ms pulse, 1,000 times in each of the X, Y, and Z directions | |
Material | Sensor main unit | PPS/PES/PBT/SUS303/SUS304/SUSXM7 |
Sensor surface | Rubber | |
Mounting bracket | SUS304/PA/SUSXM7 | |
Weight | Approx. 340 g (including mounting bracket) | |
*1 Liquid must allow for the passage of an ultrasonic pulse, as well as not contain large air pockets or excessive bubbles. Detection may be unstable on certain non-standard pipes (i.e. lined pipes) *2 Contact KEYENCE when the temperature of the pipe is greater than 85°C *3 The zero cut flow rate can be changed in the settings. When using the unit with a low flow rate range, perform an origin adjustment when the fluid is not moving if you change the zero cut flow rate. *4 This specification is valid when the flow velocity distribution is stable. This value does not take into account the effects of pulsation or fluctuations in flow velocity distribution due to facility factors. Convert the F.S. (full scale) listed in the table according to the rated flow range. *5 IO-Link: Compatible with Specification v1.1 / COM2 (38.4 kbps) The setting file can be downloaded from the KEYENCE website. If using the unit in the environment where downloading the file is not possible via internet, contact your nearest KEYENCE office. IO-Link is either registered trademarks or trademarks of PROFIBUS Nutzerorganisation e.V. (PNO) *6 When including the loads, please add 200 mA to this value. |